| Non-destructive analysis, 3D constituent density distribution analysis, 3D spectroscopic measurement | |||||||||||||
|
|||||||||||||
|
High dynamic range module : 0.02THz to 0.6THz Broadband module (*2) : 0.05THz to 3THz |
|||||||||||||
|
High dynamic range module : >60dB (at peak, 3min.) Broadband module : >50dB (at peak, 15min.) |
|||||||||||||
| <3mm (using wire phantom; modulation transfer function (MTF), 10%) |
|||||||||||||
|
Maximum waveform acquisition speed : 250Hz (Spectroscopic measurement mode) Acquisition time for CT cross section : <15min. (Measurement of 1 CT cross section is divided into 64x64) |
|||||||||||||
| <Φ310mm (12.2inch) × 310mm (12.2inch) (H) <20kg |
|||||||||||||
|
Temperature : +10°C to +30°C Relative humidity : <80% (non-condensing) |
|
| AC100V (100-120) / 200V (220-240), 850VA, 50/60Hz | |
|
Size : 1500mm (W) × 1570mm (D) × 1600mm (H) Mass : 400kg or less (*3) |
| (*1) | At temperatures of 23°C ± 5°C | |
| (*2) | With dry air option | |
| (*3) | Including analysis unit and CT test bench carrier, excluding hardware options and PC | |