The TAS7500 nondestructively visualizes the distributions of physical and chemical properties in 2D and 3D. Film thickness and density variations of tablets are a complex system that determines product performance. The user interface allows development of highly automated methods for a wide variety of measurements and analyses of samples.
The TAS7500 analyzes spectral properties of complex dosage forms based on intermolecular interactions. Rapid analysis of physicochemical properties of various dosage forms - from liquids to solids - is possible by simply exchanging measurement modules (transmission, reflection, and ATR).
Advantest has created a multifunctional and modular spectroscopic system that enables operations ranging from imaging analysis to various spectral analysis in a compact, table-top sized format. High-throughput measurement and analysis is possible with Advantestfs unique high-performance THz autosampler.

Nondestructive 2D or 3D measurement of defects by THz THz spectral measurements |
|
• Specimen: |
Medicinal tablets, pharmaceutical products (powders, liquids), other reagents, chemical substances |
• Specimen dimensions: |
Tablet defect analysis mode: weight: less than 50 g Penetration/reflection mode: ATR mode: |
• Spatial resolution: |
0.3 mm (at 2 THz) |
• minimum step size for mechanical translation stage |
0.05 mm |
• Automatic measurement |
Place up to 10 tablets in a dedicated cassette and execute automatic measurements |
|
|
• Maximum frequency range: |
0.05 - 4 THz, extension to 7 THz available as an optional feature (in transmission or reflection mode, S/N = 1) (*1) |
• S/N: |
50 dB or more (average 1024 at peak frequency) (*1) |
• Thickness measurement performance |
Measurement scope: over 30 µm; resolution: 2 µm |
| Spectrometry: max. 3.3 ms/scan; imaging measurement: within 5 min (32×32 points, average 32 times) | |
| Built-in dry air filter (external air supply required) | |
| Temperature range: +10 °C to +30 °C; relative humidity: under 80% (non-condensing conditions) |
|
| AC 100 V (100-120) / 200 V (220-240), 500 VA, 50/60 Hz (*2) | |
• Imaging measurement unit: |
dimensions: 495 (W) × 490 (D) × 380 (H) mm; mass: less than 38 kg |
• Spectroscopic measurement unit: |
dimensions: 400 (W) × 490 (D) × 380 (H) mm; mass: less than 21 kg |
| (*1) | Temperature range 23±5 °C | |
| (*2) | Excluding analysis PC | |